Explore SANWOOD HAST chambers and temperature forcing systems for SiC package moisture evaluation and electrical measurements at controlled temperatures.
In semiconductor reliability testing, silicon carbide (SiC) MOSFETs and power modules present two distinct laboratory tasks: evaluating package behavior under accelerated moisture exposure and measuring electrical parameters at controlled temperatures. A useful SiC reliability testing plan defines the conditions for each task so that changes in the device can be distinguished from changes in the measurement setup.
SANWOOD manufactures HAST Test Chambers for highly accelerated stress testing and Temperature Forcing Systems for localized thermal conditioning. These systems support selected stages of a reliability program, with equipment selection guided by package construction, measurement requirements and the intended thermal load. Complete qualification also requires the other tests specified in the device’s qualification plan.
For a non-hermetic SiC package, moisture testing starts with its encapsulation, interfaces and applicable test method. A molded discrete MOSFET and a gel-filled power module should not automatically receive the same exposure simply because both contain SiC dies.
A SANWOOD HAST Test Chamber provides elevated temperature and humidity under pressure. For suitable packages and procedures, this exposure supports investigation of moisture-related corrosion, insulation degradation and electrical changes.
Common method references include JESD22-A110 for biased HAST and JESD22-A118 for unbiased HAST. The qualification plan determines the applicable revision, exposure conditions and acceptance criteria.
The following values describe a configuration published on SANWOOD's HAST product page; they should not be treated as specifications for every available configuration.
| Parameter | Published specification |
| Temperature control range | +105°C to +135°C |
| Humidity control range | 75% to 100% RH |
| Temperature fluctuation | ±0.5°C |
| Humidity fluctuation | ±2% RH |
| Temperature uniformity | 3°C |
Select temperature, humidity and pressure as a coordinated operating condition under the governing procedure. The published 100% RH upper limit is not a prescribed setting for a method requiring non-condensing exposure. Chamber transitions and representative loading also need attention to avoid unintended condensation.
For biased HAST, the equipment review should cover the bias source, electrical feedthroughs, fixture insulation and monitoring responsibilities. Environmental performance alone does not establish high-voltage suitability. A specified high-voltage temperature–humidity reverse-bias test should not be replaced with HAST without an approved technical basis.
A SANWOOD Temperature Forcing System directs temperature-controlled air toward a selected device or test area while external instruments perform the electrical measurements.
For SiC MOSFET testing, this arrangement can support comparisons of on-state resistance, leakage and threshold voltage at selected temperatures. The fixture needs to expose the intended area to airflow while maintaining electrical connections and access for temperature sensing.
The following published specifications describe a SANWOOD temperature forcing configuration. Achievable sample temperatures and stabilization times depend on the assembly, airflow arrangement and operating load.
| Parameter | Published specification |
| Temperature control range | −80°C to +225°C |
| Typical delivered-air transition | Approximately 10 seconds between −55°C and +125°C in either direction |
| Airflow output | 4–18 scfm |
| Control modes | Air Mode and DUT Mode |
| External temperature feedback | Type-T and Type-K thermocouple inputs |
The transition figure describes the delivered air. Electrical measurements should begin after the specified sample location meets the required temperature tolerance and stabilization criterion.
Air Mode uses air temperature as the control reference. DUT Mode uses feedback from a defined point on the device under test. A thermocouple attached to the package measures that location; junction temperature requires an appropriate measurement or estimation method.
A parameter shift can reflect measurement conditions as well as device degradation. Recording the temperature reference together with the electrical settings helps the laboratory interpret the result.
| Measurement | Conditions to define | Main interpretation concern |
| On-state resistance, RDS(on) | Gate-source voltage, drain current, pulse duration or reading time, and sensing connections | Self-heating and connection resistance can affect the reading |
| Drain-source leakage, IDSS | Drain-source voltage, gate state, temperature and measurement delay | Fixture leakage can contribute to the measured current |
| Gate threshold voltage, VGS(th) | Current criterion, gate-bias history, sweep direction and readout timing | Charge trapping can affect measurement reproducibility |
For RDS(on), the measurement procedure needs to account for self-heating. A stable package-surface temperature before a current pulse does not establish that junction temperature remains unchanged during the pulse.
Threshold-voltage measurements require particular attention to electrical history. Where the adopted method specifies gate conditioning, keep the conditioning voltage, duration and delay before readout consistent. Thermal stabilization alone cannot remove differences caused by previous gate bias.
Temperature forcing provides a controlled thermal condition for this work. The laboratory’s electrical procedure determines how the parameter is measured, and additional stress-test evidence is needed to assess long-term reliability.
An assembled power module may respond very differently from a discrete device exposed to the same airflow. Busbars, mounting hardware and an active cold plate create additional heat-transfer paths, while powered operation adds device losses.
For a SANWOOD temperature forcing review, identify the area that must reach temperature and state whether the cooling circuit remains active. Assembly dimensions, expected dissipation, thermal-head access and sensor locations help establish whether localized airflow suits the intended test.
For example, a sensor close to the airflow outlet may reach its target while a module baseplate remains influenced by the coolant circuit. If temperatures at several locations matter to the measurement, monitor those locations rather than treating one feedback point as representative of the entire module.
This assessment connects equipment capability to the installed assembly and load—the conditions under which the electrical data will actually be collected.
Where the approved test plan permits, temperature forcing can support electrical measurements before and after a separate HAST exposure. The exposure and measurement setups perform different tasks within the laboratory's planned sequence.
Consider a development investigation comparing two encapsulation constructions. A post-exposure leakage difference is easier to interpret when the baseline and subsequent measurements use the same voltage, temperature reference and readout timing. Package inspection may still be needed to identify the cause.
Record sample identity, exposure history, recovery conditions and elapsed time before measurement. Additional heating or drying can change retained moisture. A post-HAST temperature sweep therefore belongs in the procedure only when it is compatible with the investigation and prescribed readout requirements.
Link the SANWOOD chamber exposure record to the electrical measurement record through sample identity. Keeping both records available helps the team investigate whether a difference is associated with exposure, recovery or measurement conditions.
For a HAST request, provide package construction, sample quantity, fixture layout, exposure conditions and bias requirements. For temperature forcing, include assembly geometry, target temperatures, sensor locations, heat dissipation and cooling status.
SANWOOD can use this information to review the proposed environmental equipment and configuration against the actual test arrangement. Discussing these details early helps identify fixture, airflow and electrical-interface requirements before equipment selection is finalized.
Contact info@sanwood.com with your sample information and test objectives to discuss a SANWOOD HAST Test Chamber or Temperature Forcing System for your SiC testing application.
Consider batch sizes, scheduling and the need for separate exposure profiles. SANWOOD offers single-chamber and vertically stacked dual-chamber configurations with independent upper and lower chamber control. Compare usable fixture capacity in each chamber alongside the available installation space.
Check supply-air dryness, ambient moisture ingress and exposed cold surfaces. Depending on the fixture, a shroud or dry-air purge may be needed to manage condensation or frost. Ask SANWOOD to confirm the required supply-air dew point and which moisture-control provisions are included in the proposed setup.
Define the temperature commands, status signals and measurement synchronization the tester requires. SANWOOD can then provide information on the supported functions and protocols for the proposed configuration. Compatibility should be established from those functions and the tester's requirements before integration.